FDA-Device2016-03-23Class II
Prime Series Stretchers (Fifth Wheel, Big Wheel and Zoom) Powered stretcher
Stryker Medical Division of Stryker Corporation
Hazard
The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we
Sold states
Distributed in DC and the states of AL, CA, FL, MN, MT, NC, NE, NY, PA, RI, TN, TX, UT, and WV.
Affected count
36
Manufactured in
3800 E Centre Ave, Portage, MI, United States
Products
Prime Series Stretchers (Fifth Wheel, Big Wheel and Zoom) Powered stretcher
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1184-2016Don't want to check this manually?