FDA-Device2016-03-30Class II
Algostim LLC, Algovita External Pulse Generator, Trial Stimulator, Model 3400
QiG Group LLC
Hazard
The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.
Sold states
Distributed in Germany and Luxembourg.
Affected count
31
Manufactured in
10675 Naples St NE, N/A, Blaine, MN, United States
Products
Algostim LLC, Algovita External Pulse Generator, Trial Stimulator, Model 3400
Official notice