FDA-Device2019-04-24Class II
Innova IGS 630, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
GE Healthcare, LLC
Hazard
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Sold states
Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.
Affected count
49 devices
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
Innova IGS 630, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Official notice