FDA-Device2014-03-19Class II

Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Mar Cor Purification
Hazard

There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

Sold states
nationwide. No distribution outside the US.
Affected count
130
Manufactured in
14550 28th Ave N, N/A, Minneapolis, MN, United States
Products
Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1201-2014