FDA-Device2026-02-04Class IIPROCESSING DEFECT
Zimmer A.T.S. 5000TS Tourniquet Systems recalled for screen freezing in non-English languages
OTHERNationwide distribution
Check your Zimmer tourniquet system software
The screen on Zimmer A.T.S. 5000TS Tourniquet Systems may freeze or stop responding when used in certain non-English language settings. This can delay surgery while staff locate an alternate device. The issue affects systems running software version v2.04 or earlier.
- Check if your A.T.S. 5000TS tourniquet (Model 60500010100) is running software v2.04 or prior
- Contact Zimmer Surgical for a software update or replacement device
- If operating in a non-English language, verify the device is functional before use
Hazard
Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
Sold states
International distribution to the countries of Canada and EMEA only.
Affected count
206
Manufactured in
200 W Ohio Ave, N/A, Dover, OH, United States
Products
Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S.¿ 5000TS Product Description: A.T.S 5000TS Tourniquet Systems Component: N/A
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1201-2026Don't want to check this manually?