FDA-Device2026-02-04Class IIPROCESSING DEFECT
Zimmer A.T.S. 3200TS Tourniquet System recalled for screen freeze in non-English languages
OTHERNationwide distribution
Check your Zimmer A.T.S. 3200TS tourniquet software
The touchscreen on certain Zimmer A.T.S. 3200TS tourniquet systems may freeze or stop responding when the device is set to operate in certain non-English languages. This could delay surgery while staff locate an alternate device. The issue affects systems running software version 2.06 or earlier.
- If your facility uses an A.T.S. 3200TS tourniquet, check the software version on your device(s)
- Contact Zimmer Surgical Inc. for a software update if your version is 2.06 or prior
- Until updated, avoid using non-English language settings on affected devices
Hazard
Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
Sold states
International distribution to the countries of Canada and EMEA only.
Affected count
51
Manufactured in
200 W Ohio Ave, N/A, Dover, OH, United States
Products
Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 3200TS Tourniquet Systems Model/Catalog Number: 60320010100 Software Version: Software version v2.06 or prior are within scope of the recall for A.T.S.¿ 3200TS Product Description: A.T.S 3200TS Tourniquet Systems Component: N/A
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1202-2026Don't want to check this manually?