FDA-Device2025-03-05Class II

A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H RT: Anatomic Lateral Fibula Plate, 8-hole, Right Indicated for use in Fixation of fractures of the distal tibia Model/Catalog Number: 77070802

Tyber Medical
Hazard

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

Sold states
Nationwide
Affected count
212 units
Manufactured in
83 S Commerce Way Ste 310, N/A, Bethlehem, PA, United States
Products
A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H RT: Anatomic Lateral Fibula Plate, 8-hole, Right Indicated for use in Fixation of fractures of the distal tibia Model/Catalog Number: 77070802

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1214-2025