FDA-Device2015-03-04Class II

The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)

Brainlab AG
Hazard

The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.

Sold states
Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
Affected count
114 offset cup impactors
Manufactured in
Kapellenstrasse 12, N/A, Feldkirchen, N/A, Germany
Products
The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1215-2015

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief