FDA-Device2017-02-22Class III
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
Siemens Healthcare Diagnostics, Inc.
Hazard
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
Sold states
Distributed throughout the United States
Affected count
13,147 units
Manufactured in
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, United States
Products
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1216-2017Don't want to check this manually?