FDA-Device2019-05-01Class II

Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.

Smiths Medical ASD Inc.
Hazard

Several lots of Jelco¿ ViaValve" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process.

Sold states
Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, and New Zealand.
Affected count
136,600 units
Manufactured in
6000 Nathan Ln N, N/A, Minneapolis, MN, United States
Products
Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4", Model Number REF 3265, MOD 10 The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1218-2019