FDA-Device2015-04-01Class II

GE Inspection Technologies x|argos 160D system. Cabinet x-ray system.

GE Inspection Technologies, LP
Hazard

It was discovered that the upper left corner-shaped lead shield in the window in one of the doors installed had lost adhesion and fall out of position, which could lead to gaps in shielding.

Sold states
US Distribution to the states of OH and PA.
Affected count
2
Manufactured in
50 Industrial Park Rd, Lewistown, PA, United States
Products
GE Inspection Technologies x|argos 160D system. Cabinet x-ray system.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1225-2015