FDA-Device2022-06-15Class II

STORZ KARL STORZ-ENDOSKOPE REF 11272VUE CMOS Video Cysto-Urethroscope REF 11272VUEK CMOS Video Cysto-Urethroscope C-VIEW KIT REF 11272VUE-R CMOS Video Cysto-Urethroscope REF 11272VUEK-R CMOS Video Cysto-Urethroscope C-VIEW KIT

Karl Storz Endoscopy
Hazard

The endoscope may experience a loss of image when activating monopolar electrocautery, due a manufacturing assembly process issue.

Sold states
Worldwide distribution - U.S. Nationwide distribution in the states of AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, VA, VT, WA, WI, WV, and WY. The country of Canada.
Affected count
401 units
Manufactured in
2151 E Grand Ave, El Segundo, CA, United States
Products
STORZ KARL STORZ-ENDOSKOPE REF 11272VUE CMOS Video Cysto-Urethroscope REF 11272VUEK CMOS Video Cysto-Urethroscope C-VIEW KIT REF 11272VUE-R CMOS Video Cysto-Urethroscope REF 11272VUEK-R CMOS Video Cysto-Urethroscope C-VIEW KIT

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1234-2022