FDA-Device2025-03-05Class II

Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation. Model: EPK-i8020c

Pentax of America Inc
Hazard

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

Sold states
Nationwide
Affected count
94 units
Manufactured in
3 Paragon Dr, N/A, Montvale, NJ, United States
Products
Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation. Model: EPK-i8020c

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1238-2025

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief