FDA-Device2019-05-08Class II
Ingenia Ambition S (781359), Ingenia Ambition X (781356) with Magnet Energization Device (MED), Philips MR systems are indicated for use as a diagnostic device. The purpose of the MED is to supply power to charge the magnet. MED is abbreviation for Magnet Energize Device
Philips Medical Systems Nederlands
Hazard
Magnet Energization Device (MED) can have a residual voltage on output terminal, may cause electrical shocks to personnel servicing the MED, MDD or connected cabling.
Sold states
US Distribution to state of: FL and Internationally to: Switzerland and Germany.
Affected count
4 units
Manufactured in
Veenpluis 4-6, N/A, Best, N/A, Netherlands
Products
Ingenia Ambition S (781359), Ingenia Ambition X (781356) with Magnet Energization Device (MED), Philips MR systems are indicated for use as a diagnostic device. The purpose of the MED is to supply power to charge the magnet. MED is abbreviation for Magnet Energize Device
Official notice