FDA-Device2016-03-30Class II
IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use the IMMULITE¿2000 systems Analyzers- for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders
Siemens Healthcare Diagnostics, Inc
Hazard
The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.
Sold states
Worldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
Affected count
9334 units
Manufactured in
333 Coney St, N/A, East Walpole, MA, United States
Products
IMMULITE 2000 EstradiolTest Code E2 Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D) SMN: 10381178, 10702833, 10381177, 10702834 Product Usage: For in vitro diagnostic use the IMMULITE¿2000 systems Analyzers- for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders