Moderate riskFDA DeviceClass II

Focalyx Fusion prostate biopsy software recalled for Windows 10 compatibility issue

Focalyx Fusion, a software device used to guide prostate biopsies and diagnostic procedures, may not work properly on Windows 10 operating systems. This malfunction could harm patients during biopsy or delay needed treatment. The manufacturer recommends stopping use of affected software versions until the device is confirmed to work with Windows 11.

From the FDA notice · Read the official notice ↗

What to do

  1. Stop using Focalyx Fusion software (versions 1.3, 1.5, or 1.6) immediately if your system runs Windows 10
  2. Contact Focalyx Technologies to verify your device serial number and get an updated version compatible with Windows 11
  3. Do not perform prostate biopsies or interventional procedures using this software until compatibility is confirmed

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Focalyx Fusion
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
18
Illnesses reported
None reported
Manufactured in
2140 W 68th St Ste 204, N/A, Hialeah, FL, United States
Classification
Class II
Recall number
Z-1243-2026

In the FDA's words

Description
Focalyx Fusion Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11. UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324 Distribution: Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO. Affected quantity: 18
Hazard
Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.