FDA-Device2016-03-30Class II
MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions, tumors. conditions anywhere in the body where radiation therapy is indicated.
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Hazard
Due to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from "Completed (or Discontinued)" to "Untreated" under specific conditions.
Sold states
Distributed in the states of Florida, New York, Ohio & Texas, and the countries of France, Germany, Japan, Italy, Korea, & Belgium.
Affected count
25 devices (4 domestically in U.S.A. and 21 internationally)
Manufactured in
6-22, 4-CHOME, KAN-ON-SHIN-MACHI, NISHI-KU, HIROSHIMA, N/A, Japan
Products
MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after) Intended for radiation therapy of lesions, tumors. conditions anywhere in the body where radiation therapy is indicated.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1244-2016Don't want to check this manually?