FDA-Device2014-03-26Class II

Natural-Knee¿ II - Unicompartmental Knee Prosthesis: CONG TIB INS SZ1/2 9MM UN CONG TIB INS SZ1/2 11MM U CONG TIB INS SZ1/2 13MM U CONG TIB INS SZ3/4 9MM UN CONG TIB INS SZ3/4 11MM U CONG TIB INS SZ3/4 13MM U CONG TIB INS SZ 5/6 9MM U CONG TIB INS SZ5/6 11MM U CONG TIB INS SZ5/6 13MM U

Zimmer, Inc.
Hazard

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Sold states
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Affected count
3638 units
Manufactured in
1800 W Center St, Warsaw, IN, United States
Products
Natural-Knee¿ II - Unicompartmental Knee Prosthesis: CONG TIB INS SZ1/2 9MM UN CONG TIB INS SZ1/2 11MM U CONG TIB INS SZ1/2 13MM U CONG TIB INS SZ3/4 9MM UN CONG TIB INS SZ3/4 11MM U CONG TIB INS SZ3/4 13MM U CONG TIB INS SZ 5/6 9MM U CONG TIB INS SZ5/6 11MM U CONG TIB INS SZ5/6 13MM U

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1251-2014