FDA-Device2014-03-26Class II

Natural-Knee¿/Apollo" Knee: M/B PAT SZ0 NATURAL-KNEE M/B PAT SZ1 NATURAL-KNEE M/B PAT SZ2 NATURAL-KNEE

Zimmer, Inc.
Hazard

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Sold states
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Affected count
3638 units
Manufactured in
1800 W Center St, Warsaw, IN, United States
Products
Natural-Knee¿/Apollo" Knee: M/B PAT SZ0 NATURAL-KNEE M/B PAT SZ1 NATURAL-KNEE M/B PAT SZ2 NATURAL-KNEE