FDA-Device2014-03-26Class II
Natural-Knee¿ II System ULTRA CONG INS LT SZA/9MM ULTRA CONG INS LT SZA/13M ULTRA CONG INS LT SZA/19M ULTRA CONG INS RT SZA/9MM ULTRA CONG INS RT SZA/11M ULTRA CONG INS RT SZA/13M ULTRA CONG INS RT SZA/19M ULTRA CONG INS LT SZB/13M ULTRA CONG INS LT SZB/16M ULTRA CONG INS RT SZB/19M ULTRA CONG INS RT SZB/22M NKII ROT PLTFM U/C INS, 9 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U
Zimmer, Inc.
Hazard
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
Sold states
Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Affected count
3638 units
Manufactured in
1800 W Center St, Warsaw, IN, United States
Products
Natural-Knee¿ II System ULTRA CONG INS LT SZA/9MM ULTRA CONG INS LT SZA/13M ULTRA CONG INS LT SZA/19M ULTRA CONG INS RT SZA/9MM ULTRA CONG INS RT SZA/11M ULTRA CONG INS RT SZA/13M ULTRA CONG INS RT SZA/19M ULTRA CONG INS LT SZB/13M ULTRA CONG INS LT SZB/16M ULTRA CONG INS RT SZB/19M ULTRA CONG INS RT SZB/22M NKII ROT PLTFM U/C INS, 9 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 NKII ROT PLTFM U/C INS, 1 ALL POLY CONG TIB LT SZ00 ALL POLY CONG TIB LT SZ0/ ALL POLY CONG TIB LT SZ0/ ALL POLY CONG TIB RT SZ0/ ALL POLY CONG TIB RT SZ0/ ALL POLY CONG TIB LT SZ1/ ALL POLY CONG TIB LT SZ1/ ALL POLY CONG TIB RT SZ1/ ALL POLY CONG TIB RT SZ1/ ALL POLY CONG TIB RT SZ1/ ALL POLY CONG TIB LT SZ2/ ALL POLY CONG TIB LT SZ2/ ALL POLY CONG TIB LT SZ2/ ALL POLY CONG TIB LT SZ2/ ALL POLY CONG TIB RT SZ2/ ALL POLY CONG TIB RT SZ2/ ALL POLY CONG TIB RT SZ2/ ALL POLY CONG TIB LT SZ3/ ALL POLY CONG TIB LT SZ3/ ALL POLY CONG TIB LT SZ3/ ALL POLY CONG TIB RT SZ3/ ALL POLY CONG TIB RT SZ3/ ALL POLY CONG TIB RT SZ3/ ALL POLY CONG TIB LT SZ4/ ALL POLY CONG TIB LT SZ4/ ALL POLY CONG TIB RT SZ4/ ALL POLY CONG TIB RT SZ4/ ALL POLY CONG TIB RT SZ5/ PS TIB INS RT SZ1/2 16MM PS TIB INS RT SZ1/2 19MM PS TIB INS LT SZ00/0 19MM PS TIB INS RT SZ00/0 11MM PS TIB INS RT SZ00/0 16MM
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1255-2014Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief