FDA-Device2022-06-15Class II

Reliance Synergy Washer/Disinfector

Steris Corporation
Hazard

The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

Sold states
Worldwide distribution - US Nationwide.
Affected count
3,863 total units (2,776 US; 1,087 OUS)
Manufactured in
5960 Heisley Rd, Mentor, OH, United States
Products
Reliance Synergy Washer/Disinfector

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1256-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief