FDA-Device2022-06-15Class II
Reliance Synergy Washer/Disinfector
Steris Corporation
Hazard
The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Sold states
Worldwide distribution - US Nationwide.
Affected count
3,863 total units (2,776 US; 1,087 OUS)
Manufactured in
5960 Heisley Rd, Mentor, OH, United States
Products
Reliance Synergy Washer/Disinfector
Official notice