FDA-Device2023-03-22Class II

Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Model: 650705550001

Stryker Medical Division of Stryker Corporation
Hazard

Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.

Sold states
US Nationwide distribution in the states of NY, VA.
Affected count
4 units
Manufactured in
3800 E Centre Ave, N/A, Portage, MI, United States
Products
Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Model: 650705550001

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1256-2023