FDA-Device2022-06-22Class II

Olympus CYSTONEPHROFIBERSCOPE Model: CYF-5

Hazard

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Sold states
CA FL MA MD MI NJ OH PR WI
Affected count
7 units
Manufactured in
3500 Corporate Pkwy, Center Valley, PA, United States
Products
Olympus CYSTONEPHROFIBERSCOPE Model: CYF-5

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1262-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief