FDA-Device2014-03-26Class II
Philips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.
Philips Medical Systems North America Inc.
Hazard
A problem has been detected with Philips Model EPIQ 5 and 7 Ultrasound System. The system will report incorrect Doppler velocities if you use High Pulse Repetition Frequency (HPRF) with multiple sample volume gates. While in High Pulse Repetition Frequency (HPRF) in Display Zoom Velocity Scale using manual spectral Doppler trace, the velocities in the velocity results table may be overstated b
Sold states
Worldwide Distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WV; and countries of: Australia, Austria, Belgium, Canada, Chili, Ecuador, Egypt, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Korea Rep, Malaysia, Netherlands, New Zealand, Norway, Peru, Philippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.
Affected count
1267 total :670 units in US and 1267 units Internationally (which includes 11 units in Canada)
Manufactured in
22100 Bothell Everett Hwy, Bothell, WA, United States
Products
Philips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1264-2014Don't want to check this manually?