FDA-Device2021-03-24Class II
Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100
Imactis
Hazard
Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory
Sold states
NJ, NY, WI
Affected count
105 kits
Manufactured in
20 Rue Du Tour De L Eau, N/A, St Martin D Heres, N/A, France
Products
Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100