FDA-Device2021-03-24Class II

Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100

Imactis
Hazard

Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory

Sold states
NJ, NY, WI
Affected count
105 kits
Manufactured in
20 Rue Du Tour De L Eau, N/A, St Martin D Heres, N/A, France
Products
Impactis NaviKit-stereotaxic accessory for Computed Tomography systems Reference # I10100

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1264-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief