FDA-Device2025-03-12Class II

MIM software; System, Image Processing, Radiological

MIM Software Inc
Hazard

In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.

Sold states
Worldwide distribution - United States Nationwide and the countries of Australia, Belgium, Bolivia, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Israel, Italy, Japan, Mexico, Netherlands, Poland, Russia, Spain, Sweden, United Kingdom, Finland, Hong Kong.
Affected count
352 units
Manufactured in
25800 Science Park Dr Ste 180, Cleveland, OH, United States
Products
MIM software; System, Image Processing, Radiological

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1267-2025

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief