FDA-Device2014-04-02Class III

Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer

Lin-Zhi International Inc
Hazard

Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).

Sold states
US Distribution including CA.
Affected count
120 units, 60 of each lot number
Manufactured in
670 Almanor Ave, N/A, Sunnyvale, CA, United States
Products
Single Analyte DAU Calibrator (5 mL) Norbuprenorphine: Cutoff Calibrator; IVD - for in vitro diagnostic use only; Lin-Zhi International Sunnyvale, CA The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1270-2014