FDA-Device2025-03-12Class II

Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;

Telcare, LLC
Hazard

Telcare, LLC has received customer complaints related to battery expansion/swelling for its Philips Connected Blood Glucose Meter (i.e., BGM 4 brand name) Monitoring System.. The issue was identified by post market surveillance data (customer complaints) - events of battery expansion/swelling and associated symptoms, e.g., leading to the device's removable back cover (battery port) to separate. If there is a loss of primary function of the device due to the described device defect, an immediate health consequence that may result is prolonged hypoglycemia or hyperglycemia due to delay in treatment, as the blood glucose measurement is not available to guide treatment. In a worst-case situation, there is potential for battery swelling to result in explosion of the battery and/ or device. Exposure to explosion, and subsequently projectiles from the explosion, may result in several immediate health consequences, including concussion, headache, dizziness, eye irritation, blurred vision, eye injury, and vision impairment.

Sold states
US Nationwide distribution.
Affected count
34,764 units
Manufactured in
150 Baker Avenue Ext, Ste 300, Concord, MA, United States
Products
Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1270-2025

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.