FDA-Device2016-04-06Class II

system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Siemens Medical Solutions USA, Inc
Hazard

In Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. In Artis systems with A100Plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the X-ray tube) can result in the failure of a module in the high-voltage generator.

Sold states
US Nationwide Distribution including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and PUERTO RICO.
Affected count
552 units
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1299-2016