FDA-Device2018-04-11Class II
Syngo.via. Medical Device Software. Picture archiving and communications system.
Hazard
A software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated information could be sent to the referring physician when creating the formal report in the reporting system.
Sold states
Device is software only. No products are distributed to wholesale dealers, distributers or retailers
Affected count
236 units
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Syngo.via. Medical Device Software. Picture archiving and communications system.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1303-2018More Siemens Medical Solutions USA, Inc recalls
- FDA-Device2026-07-08Siemens ARTIS icono ceiling monitor suspension recalled for loose screw risk
- FDA-Device2026-07-08Siemens ARTIS icono floor recalled for potential movement loss at detector height limits
- FDA-Device2026-07-08Siemens ARTIS icono biplane recalled for potential loss of movement after detector adjustment
- FDA-Device2026-07-08Siemens ARTIS icono ceiling model 11328100 recalled for loss of movement capability