FDA-Device2018-04-11Class II

Dynavox T15 Speech Generating Device ( including Batteries)

Dynavox Systems Llc
Hazard

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Sold states
Nationally
Affected count
307
Manufactured in
2100 Wharton St Ste Ste 400, N/A, Pittsburgh, PA, United States
Products
Dynavox T15 Speech Generating Device ( including Batteries)

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1305-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.