FDA-Device2022-07-06Class II

NaviCare Nurse Call/Voalte Nurse Call

Baxter Healthcare Corporation
Hazard

An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.

Sold states
nationwide and Canada
Affected count
283 installations
Manufactured in
1 Baxter Pkwy, Deerfield, IL, United States
Products
NaviCare Nurse Call/Voalte Nurse Call

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1306-2022