FDA-Device2015-04-01Class II

RayStation Radiation Therapy Treatment Planning System; -- RayStation 3.5, RayStation 4.0, RayStation 4.5 and RayStation 4.7. For RayStation 4.7, the issue applies also to machines with fixed jaws, regardless of MLC/jaw position. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

RAYSEARCH LABORATORIES AB
Hazard

An issue with photon dose calculation for DMLC (Dynamic MLC) plans for machines where the MLC is positioned above the jaws, e.g. some Elekta linacs. The magnitude of the error depends on the beam model output factor corrections and on the individual DMLC plan characteristics.

Sold states
Nationwide Distribution.
Affected count
Domestic: 245 units
Manufactured in
Sveavaegen 9, Stockholm, Sweden
Products
RayStation Radiation Therapy Treatment Planning System; -- RayStation 3.5, RayStation 4.0, RayStation 4.5 and RayStation 4.7. For RayStation 4.7, the issue applies also to machines with fixed jaws, regardless of MLC/jaw position. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1310-2015