FDA-Device2014-04-09Class II

Omni Micro-electrode/reference electrode for cobas b221 analyzer, Model/Catalog/Part Number: 03111873180 as a part of the following systems: 1. 03337103001, cobas b 221<1>Roche OMNI S1 system 2. 03337111001, cobas b 221<2>Roche OMNI S2 system 3. 03337154001,cobas b 221<6>Roche OMNI S6 system 4. 03337138001, cobas b 221<4>Roche OMNI S4 system Product Usage: Usage: Blood Gas and Electrolytes Analyzer c. Classification Name: Fully automated Critical Care Analyzer for the measuremen

Roche Diagnostics Operations, Inc.
Hazard

Reference electrode used beyond the guaranteed in-use 52 week lifetime,may leak and potentially cause erroneous pH and/or sodium results. Other Ion parameters are not affected.

Sold states
Worldwide Distribution - USA nationwide including PR and the country of Guam.
Affected count
N/A
Manufactured in
9115 Hague Rd, N/A, Indianapolis, IN, United States
Products
Omni Micro-electrode/reference electrode for cobas b221 analyzer, Model/Catalog/Part Number: 03111873180 as a part of the following systems: 1. 03337103001, cobas b 221<1>Roche OMNI S1 system 2. 03337111001, cobas b 221<2>Roche OMNI S2 system 3. 03337154001,cobas b 221<6>Roche OMNI S6 system 4. 03337138001, cobas b 221<4>Roche OMNI S4 system Product Usage: Usage: Blood Gas and Electrolytes Analyzer c. Classification Name: Fully automated Critical Care Analyzer for the measurement of pH, Blood gases, electrolytes, Hematocrit, hemoglobin, glucose, lactate, urea/BUN, total hemoglobin, Oxygen saturation, oxy - deoxycarboxy and methemoglobin

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1315-2014

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.