FDA-Device2019-05-22Class II
SureSigns VS3 NBP, SpO2, Rec, Product Number 863072. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
Philips North America, LLC
Hazard
System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Sold states
Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.
Affected count
3177
Manufactured in
3000 Minuteman Rd, Andover, MA, United States
Products
SureSigns VS3 NBP, SpO2, Rec, Product Number 863072. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1320-2019Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief