FDA-Device2015-04-01Class II
Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004, Non-Sterile, Manufactured by Titan Spine, LLC. Endo TCS driver is one of the instruments within the surgical tray for use with the Interbody Fusion Device(IBD).
TITAN SPINE, LLC
Hazard
The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.
Sold states
US Distribution to the states of : AZ, CA, CT, FL, GA, ID, IL, KY, MD, MO, NV, NY, OH, OK, OR, TX, UT and WI.
Affected count
26
Manufactured in
6140 W Executive Dr, Suite A, Mequon, WI, United States
Products
Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004, Non-Sterile, Manufactured by Titan Spine, LLC. Endo TCS driver is one of the instruments within the surgical tray for use with the Interbody Fusion Device(IBD).
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1324-2015Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief