FDA-Device2016-04-13Class II
Fresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System Fresenius 2008K@home Hemodialysis Machine with bibag System 2008T Hemodialysis Machine The Fresenius 2008¿ Series Machines are indicated for acute and chronic dialysis therapy
Fresenius Medical Care Renal Therapies Group, LLC
Hazard
Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
Sold states
Nationwide Distribution
Affected count
8384 units (8369 US) (15 OUS)
Manufactured in
920 Winter St, N/A, Waltham, MA, United States
Products
Fresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System Fresenius 2008K@home Hemodialysis Machine with bibag System 2008T Hemodialysis Machine The Fresenius 2008¿ Series Machines are indicated for acute and chronic dialysis therapy
Official notice