FDA-Device2025-03-19Class II

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

Beckman Coulter, Inc.
Hazard

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results

Sold states
US Nationwide. Global Distribution.
Affected count
245 units
Manufactured in
1000 Lake Hazeltine Dr, Chaska, MN, United States
Products
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1326-2025

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief