FDA-Device2020-02-26Class II

CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.

Philips Medical Systems Gmbh, DMC
Hazard

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Sold states
Worldwide distribution  US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Affected count
116
Manufactured in
Development And Manufacturing Ctr., Rontgenstr. 24, Hamburg, Germany
Products
CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1328-2020

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief