FDA-Device2021-04-07Class II

Product Label - ACUSON Sequoia *** Power Input: 100-240V~, 13.0-5.4A, 50-60Hz *** (240)11148775 Model User Manual states - ACUSON Sequoia Diagnostic Ultrasound Systems

Siemens Medical Solutions USA, Inc.
Hazard

The ultrasound system averages the Mean Sac Diameter (MSD) and Gestational Sac Diameter (GSD) into the Estimated Date of Delivery (EDD) calculation. This may result in an incorrect EDD, which may influence patient management decisions regarding induction of labor and elective caesarean delivery, which may result in premature births.

Sold states
U.S.: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY O.U.S.: U.A.E., Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Morocco, Netherlands, New Zealand, Oman, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, and Vietnam
Affected count
1,172 systems
Manufactured in
22010 SE 51st St, Issaquah, WA, United States
Products
Product Label - ACUSON Sequoia *** Power Input: 100-240V~, 13.0-5.4A, 50-60Hz *** (240)11148775 Model User Manual states - ACUSON Sequoia Diagnostic Ultrasound Systems

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1330-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief