FDA-Device2021-04-14Class I

Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 1440

Heartware, Inc.
Hazard

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

Sold states
Worldwide
Affected count
17992 units (8054 US)
Manufactured in
14400 Nw 60th Ave, N/A, Miami Lakes, FL, United States
Products
Power Supply: 1. Medtronic HVAD Controller AC Adapter, Product Numbers: 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US 2. Medtronic HVAD Controller DC Adapter, Product Number: 1440

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1337-2021