Serious riskFDA DeviceClass II

Elekta Leksell GammaPlan software recalled for treatment planning error risk

Elekta's Leksell GammaPlan treatment planning software (versions 11.3 and 11.4) has a software error that could cause incorrect treatment location data if users do not properly update reference settings. This could result in radiation therapy being directed to the wrong location if the error goes undetected.

From the FDA notice · Read the official notice ↗ · All Elekta recalls

What to do

  1. If you operate or manage a Gamma Knife facility, contact Elekta immediately about this issue
  2. Review recent treatment plans created with versions 11.3 or 11.4 to ensure correct patient locations were targeted
  3. Install any software update or patch from Elekta when available
  4. Ensure staff verify that obsolete reference settings are properly changed before creating new treatment plans

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 2 products — see all of them.

Details

Units affected
N/A
Illnesses reported
None reported
Manufactured in
400 Perimeter Center Ter NE Ste 50, N/A, Atlanta, GA, United States
Classification
Class II
Recall number
Z-1340-2026

In the FDA's words

Description
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning. Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users. Lot Code: 07340048311878 (LGP 11.3) 07340201502136 (LGP 11.4) Distribution: Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam. Affected quantity: N/A
Hazard
Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.