Elekta Leksell GammaPlan software recalled for treatment planning error risk
Contact Elekta about Leksell GammaPlan software update
Elekta's Leksell GammaPlan treatment planning software (versions 11.3 and 11.4) has a software error that could cause incorrect treatment location data if users do not properly update reference settings. This could result in radiation therapy being directed to the wrong location if the error goes undetected.
- If you operate or manage a Gamma Knife facility, contact Elekta immediately about this issue
- Review recent treatment plans created with versions 11.3 or 11.4 to ensure correct patient locations were targeted
- Install any software update or patch from Elekta when available
- Ensure staff verify that obsolete reference settings are properly changed before creating new treatment plans
Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.