FDA-Device2022-07-13Class II

Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444905/RF421C

Siemens Healthcare Diagnostics, Inc.
Hazard

Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

Sold states
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.
Affected count
113 units; Extended: 52 units
Manufactured in
500 Gbc Dr, N/A, Newark, DE, United States
Products
Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444905/RF421C

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1341-2022