FDA-Device2022-07-13Class II

Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc
Hazard

It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.

Sold states
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, FL, IL, LA, ME, MN, MO, NC, OH, PA, SC, TN, TX, WI and the countries of Australia, Austria, Azerbaijan, Belgium, Canada, China, Finland, France, Germany, Hong Kong, India, Italy, Japan, Lebanon, Lithuania, Luxembourg, Netherlands, Poland, Portugal, South Korea, Switzerland, United Kingdom.
Affected count
233 worldwide; 32 US
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1343-2022