FDA-Device2023-04-12Class II

Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 18X4,1 BLA,1PRT QUICK Catalog number: 5921-218-135NS

Stryker Corporation
Hazard

Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications

Sold states
US Nationwide distribution.
Affected count
1730 units (packs of 10)
Manufactured in
1941 Stryker Way, Portage, MI, United States
Products
Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 18X4,1 BLA,1PRT QUICK Catalog number: 5921-218-135NS

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1343-2023