FDA-Device2020-03-04Class II
COHERENCE Oncologist, Model Nos. 07333680 & 07351898
Hazard
A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.
Sold states
Domestic distribution to CA, FL, GA, IA, MA, MI, NE, NJ, NV, NY, OH, OK, PA, SD, TN, WV. Worldwide foreign distribution.
Affected count
64 Worldwide/26 Domestic
Manufactured in
40 Liberty Blvd, N/A, Malvern, PA, United States
Products
COHERENCE Oncologist, Model Nos. 07333680 & 07351898
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1346-2020More Siemens Medical Solutions USA, Inc recalls
- FDA-Device2026-07-08Siemens ARTIS icono ceiling monitor suspension recalled for loose screw risk
- FDA-Device2026-07-08Siemens ARTIS icono floor recalled for potential movement loss at detector height limits
- FDA-Device2026-07-08Siemens ARTIS icono biplane recalled for potential loss of movement after detector adjustment
- FDA-Device2026-07-08Siemens ARTIS icono ceiling model 11328100 recalled for loss of movement capability