Philips Azurion 7 B20 imaging system recalled for unexpected table movement
Philips medical imaging systems (Azurion 7 B20) may move unexpectedly when the Reset Geometry button is pressed, even when the table lock is supposed to be engaged. This affects certain software versions across three model numbers distributed worldwide.
From the FDA notice · Read the official notice ↗ · All Philips recalls
What to do
- Check if your facility has an affected Azurion 7 B20 system (models 722068, 722226, or 722236)
- Do not use the Reset Geometry button until you receive guidance from Philips or the FDA
- Contact Philips Medical Systems for instructions on a software update or corrective action
- Report any incidents of unexpected table movement to the FDA MedWatch program
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 8 products — see all of them.
Details
- Units affected
- 1503 units (326 US, 1177 OUS)
- Illnesses reported
- None reported
- Manufactured in
- Veenpluis 4-6, N/A, Best, N/A, Netherlands
- Classification
- Class II
- Recall number
- Z-1347-2026
In the FDA's words
- Description
- Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active. System Model Numbers: (1)722068, (2)722226, (3)722236; UDI-DIs: (1)884838085367, (2)884838099272, (3)884838116801; Serial Numbers: All; Distribution: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, FrenchPolynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen. Affected quantity: 1503 units (326 US, 1177 OUS)
- Hazard
- Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
8 products in this recall
Philips Azurion systems recalled for unexpected table movement
Philips is recalling eight Azurion 3, 5, and 7 X-ray system models. Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed, even when a table lock is active. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Azurion 3 M12 System Model Numbers: (1) 722063, (2) 722221, (3) 722229; Software Version Number - All versions: R1.X, R2.X, R3.XZ-1344-2026 · 291 units (27 US, 264 OUS)
- Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.XZ-1345-2026 · 802 units (45 US, 757 OUS)
- Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.XZ-1346-2026 · 655 units (106 US, 549 OUS)
- Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X (this page)Z-1347-2026 · 1503 units (326 US, 1177 OUS)
- Azurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - All versions: R1.X, R2.X, R3.XZ-1348-2026 · 1979 units (365 US, 1614 OUS)
- Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.XZ-1349-2026 · 5136 units (1743 US, 3393 OUS)
- Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.XZ-1350-2026 · 326 units (10 US, 316 OUS)
- Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.XZ-1351-2026 · 922 units (101 US, 821 OUS)
More from Philips
See all- Philips Allura Xper systems recalled for FlexVision software issue
ModerateFDANov 6, 2026
- Philips Azurion 3 M12 systems recalled over unapproved drives
ModerateFDAOct 27, 2026
- Philips Azurion X-ray systems recalled over image and movement software issues
ModerateFDAAug 19, 2026
- Philips Azurion X-ray systems recalled for software image and movement issues
ModerateFDAAug 19, 2026
- Philips Allura Xper FD X-ray systems recalled over firmware issue
ModerateFDAAug 19, 2026