FDA-Device2026-02-18Class IIPROCESSING DEFECT
Philips Azurion 7 M20 imaging system recalled for unexpected table movement
OTHERNationwide distribution
Stop using affected Philips Azurion 7 M20 systems
Philips Azurion 7 M20 imaging systems may move unexpectedly when the Reset Geometry button is pressed, even if the table lock is engaged. This can occur under certain conditions across all software versions (R1.X, R2.X, R3.X). No injuries have been reported.
- Stop using the Reset Geometry button on affected Azurion 7 M20 systems until you receive further instructions from Philips
- Contact Philips Medical Systems for guidance on software updates or device modifications
- Verify your system's model number (722079, 722224, 722234, or 722282) against the recall notice
Hazard
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Sold states
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, FrenchPolynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
Affected count
5136 units (1743 US, 3393 OUS)
Manufactured in
Veenpluis 4-6, N/A, Best, N/A, Netherlands
Products
Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1349-2026More PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls
- FDA-Device2026-06-03Philips Azurion imaging systems recalled for hard drive failure risk affecting imaging functionality
- FDA-Device2026-06-03Philips Allura medical imaging systems recalled for hard drive failure risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for cooling hose degradation risk
- FDA-Device2026-05-27Philips Integris-Allura X-ray systems recalled for cooling unit oil leakage risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for impaired table movement due to control wear