FDA-Device2019-05-29Class II
VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest
Hazard
Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported
Sold states
Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI Foreign: Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines
Affected count
85 units: US=19;OUS=66
Manufactured in
100 Indigo Creek Dr, N/A, Rochester, NY, United States
Products
VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1364-2019Don't want to check this manually?