FDA-Device2018-04-18Class II
Liberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usage: The device is indicated for acute and chronic peritoneal dialysis.
Fresenius Medical Care Renal Therapies Group, LLC
Hazard
The recalling firm identified a software issue related to the Patient Line Check (PLC) which may result in an increased risk of Overfill (also known as Increased Intraperitoneal Volume, IIPV). Overfill/IIPV may result in serious injury or death.
Sold states
Nationwide distribution.
Affected count
9293
Manufactured in
920 Winter St, Waltham, MA, United States
Products
Liberty Select Cycler (SW v.2.8.7), Material Number RTLR108343 Product Usage: The device is indicated for acute and chronic peritoneal dialysis.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1365-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.